The energy-saving KIRSCH blood plasma freezer complies with DIN ISO 13485 and DIN 58375 and is equipped with the PRO-ACTIVE control system which, in addition to permanent monitoring of all performance data, ensures minimal temperature deviations between the warmest and coolest storage locations. At the same time, efficient recirculating air cooling ensures even temperature regulation without strong temperature fluctuations. This ensures constant and uniform cooling of the blood plasma samples.
To protect the temperature control from unauthorised access, it is protected with a key switch or RFID/password, depending on the version. The time-consuming documentation of the minimum and maximum temperature is minimised for the specialist staff, as the freezer is equipped with an integrated memory whose data can be easily read and archived. In the event of a power failure or other malfunctions that may cause deviations in cooling, an acoustic and visual signal indicates the malfunctions. If required, alarm forwarding to a mobile phone, for example, is also possible to receive notification of alarms anywhere and anytime.
| PRO-ACTIVE | ULTIMATE | |
| Display | Membrane keypad with digital temperature display and min./max. temperature memory. | Capacitive touch display in elegant glass design. Graphic display of the temperature curve and the min./max. temperature. |
| Temperature Regulation | Protected by key switch. | Protected by RFID/password. |
| Alarm Forwarding | Via potential-free contact (e.g. on mobile phone with GSM module or connection to central control technology). | Via network and potential-free contact (e.g. connection to central control technology). |
| Data Documentation | Reading via USB interface by means of USB stick and software KIRSCH Datanet | Digital - can be read via network (PC-KIT-NET) and KIRSCH Datanet software. |
| Access Control | No. | Yes - with protocol function. |
| Warning Function | Visual and acoustic alarm signal. | Visual and acoustic alarm signal. In the event of a power failure, the monitoring unit remains in operation for approx. 30 h thanks to the rechargeable battery. |
| BL 530 | BL 730 | |
| Capacity | 500 litres | 700 L |
| Temperature Setting | -32 °C | |
| Power Input | 740 watts | 790 watts |
| Voltage | 220-240 V, 50 Hz | |
| Standard Consumption | 10.81 kWh/24 h |
11.1 kWh/24 h |
| Permissible Ambient Temperature | +10 °C to +32 °C | |
| Heat Emission(max.) | 827 watts | 860 watts |
| External Dimensions (incl. wall distance & door handle) | 77 x 76 x 196 cm (W x D x H)* | 77 x 98 x 196 cm (W x D x H)* |
| Inner Dimensions | 62 x 56 x 140 cm (W x D x H)* | 62 x 77 x 140 cm (W x D x H)* |
| Outer Dimensions with Door Open at 90° | 77 x 144 cm (W x D) | 77 x 166 cm (W x D) |
| Light Drawer Dimensions | 56 x 39 x 10 cm (W x D x H) | 56 x 60 x 10 cm (W x D x H) |
| Max. Load Drawer | 50 kg | |
| Weight | 161 kg (net); 183 kg (gross) | 202 kg (net); 237 kg (gross) |
*Values may vary minimally depending on the variant (see product data sheet)